SPDIUV UV Control 4CT Radiometer

SPDIUV UV Control 4CT Radiometer - UV/Temperature Measuring Device
Free Gift

Brand: SPDI

SKU: 302082

UV Bulb Price Match Guarantee
$11,900.00

  • While on the conveyer belt, the UV Control can withstand max. 230° F / 110° C for up to 10 seconds. The temperature of the housing should not exceed 113° F / 45° C.

    Because of uneven radiation distribution of the UV light source and different type of construction of the measuring devices by different manufacturers, different readings may appear under the same measurement conditions.

    Calibration: In order to keep its full function and precision, it is recommended to have re-calibration done once per year. Re-calibration will also be necessary after change of battery. PTB traceable calibration with certificate.

    Attention:
    1. Please avoid shaking the UV Control.
    2. Do not expose to excessive heat.
    3. UV-light is hazardous to your health. Avoid direct UV-light to your eyes and to your body.
    Warranty: 1 year from the date of purchase.

    Control 4TC Radiometer. Features:

    • UV-A intensity mW/cm² + UV-A dose mJ/cm²
    • UV-B intensity mW/cm² + UV-B dose mJ/cm²
    • UV-C intensity mW/cm² + UV-C dose mJ/cm²
    • UV-VIS intensity mW/cm² + UV-VIS dose mJ/cm²
    • Full UV intensity mW/cm² + Full UV dose mJ/cm²
    • Temperature °C / °F
    • Permanent or triggered mode
    • SD memory card
    • Graphical and numerical display on a PC
    • Self-replaceable button cells

    The UV-Integrator UV is a self-contained, high quality UV measuring instrument. It is designed to measure, record and display peak UV intensity, UV dosage and temperature in the UV curing process.

    The instrument is equipped with four different UV sensors as well as a temperature sensor for the individual measurement of:
    • UV-C: 230 - 280 nm
    • UV-B: 280 - 315 nm
    • UV-A: 315 - 410 nm
    • UV-VIS: 395 - 445 nm
    • Temp: 32 - 230° F / 0 - 110° C

    With these four different UV bands plus the total UV band and an extra temperature measuring, most of the measuring requirements of UV curing applications can be covered. Due to its four different UV sensors and the integrated microprocessor, UV Integrator UV Control can measure, record and display the peak of the UV-intensity (mW/cm²) for each UV band individually plus the peak of total UV energy. Additionally, this UV Integrator is calculating the UV dosage (mJ/cm²) of the UV energy supplied during the time of exposure of one measuring cycle. The UV dosage is calculated for each UV band (UV-A, UV-B and UV-C) individually and as total Integral of UV-dosage over all three UV bands. This allows to determine not only the total energy, but also how that energy is delivered, i.e., what intensity and dose at what UV band.

    An extra sensor measures temperatures from 0 to 230° F / 0 to 110° C. The Integrator UV Control features a selectable "triggered mode", i.e. the recording of the measuring starts first if the incident UV intensity exceeds 2 mW/cm².

    The four sensors are on the back of the unit which also serves as a heat shield. After completion of the measuring cycle all measuring results can be scrolled through on the built in 2 x 16 digit LCD display. A special AUTO-OFF feature that turns off the unit automatically after one minute serves as energy saving and extension of the battery service life.

    This microprocessor integrator is additionally equipped with a Card Slot for the use of SD-Memory Cards. All measuring data of a measuring cycle are stored to the SD-Memory card with an identifying file name. The number of storable measuring files depends on the capacity of SD-Memory Card. Data can be loaded to a PC for further editing. The special evaluation software allows to show, edit and store a history of the measuring results of the entire measuring cycle as graphic charts (mW/cm²), (mJ/cm²) and (°C/°F).

    • Art. No.: A003472
    • Display: 2 x 16 digit LCD
    • Measuring range: 0 - 2,000 mW/cm²
    • Display range: 0 - 60,000 mJ/cm²
    • Spectral range:
    1. UV-A: 315 - 410 nm
    2. UV-B: 280 - 315 nm
    3. UV-C: 230 - 280 nm
    4. UV-V: 395 - 445 nm
    5. Full UV: 230 - 410 nm
    • Max. Power Input: 0 - 5,000 mW/cm²
    • Sampling Rate: 0.005 seconds (200/second)
    • Operating temperature: 32° - 113° F (0 to 45° C)
    • Temperature Range: 32 - 230° F (0 to 110° C)
    • Recording readiness time: 120 seconds
    • Recording cycle: 90 seconds
    • Power Supply: (2) 3.6V LiPo Akku TYP LIR 2450
    • Power Consumption: 20 uA
    • Battery service life: 1,000 charging cycles
    • Scope of delivery: UV Control Integrator, plastic case, (2) battery cells, battery charger, PC software
    • Dimensions: 140 x 65 x 12 mm
    • Weight: 17.5 ounces (500 g)
    • Heat protection: Heat shield on back plate
    • Base Accuracy: ± 5%
    • Calibration: Conforms to DIN EN ISO / IEC 17025; can be traced back to PTB
  • Display, editing and storing of measuring data on a computer:

    The graphical display of measuring data can now be further edited on a computer and the display can be further optimized to gain the desired information. This optimization is done by various function buttons. The following functions are available:

    Y-Scale/Autoscale: Adjusting the graphical display automatically to fit the screen size.

    Hairline Cursor: For the numerical display of the UV-Intensity at a certain point of a trend curve.

    Zoom In/Zoom Out: Manual enlargement of the selected area.

    Save: Saves measuring data files to a desired location.

    Copy/Transfer: Copying measuring data to export file.

    Page set-up: For the use-defined set-up of a page.

    Print preview: Shows the print preview for the selected area.

    Print: Prints the graphically displayed area.

    Change °C/°F: Allows the change of display of the temperature trend curve either in °C or °F.

    File name: Allows the allocation of details to a defined data file.

    File Import: Two data files can be loaded in one graph for comparison and evaluation.

    Synchronization: Two loaded files in one graph are synchronized for optimal comparison and evaluation.

    Fat: Enables to change the display of lines fat or soft.

    CLR: Removes a data file from the screen.

 